Model No.: AT6000
Production Capacity: 6000 units one year
Delivery Date: 3-5days
Name: Dynamic ECG Systems
ECG lead wires: 12-Lead
Recording time: 48h
User manual: 1
Net Weight: 61.5g (Without battery)
Means of Transport: Land,Air
Packing: 8.05cm(L) *5.95cm(W) *2.2cm...
Brand Name: Autonola
Certificates: CE/FDA
Sampling accuracy: 12bit
ECG electrode: 20 per bag
Warranty: 2 year
Packed Weight: about 530.00

Dynamic ECG Systems contains recorder & analysis software. It is applicable for use in medical institution and community.

Patient monitor AT6000 Dynamic ECG Systems


Recorder features

 

  •  Compact design.

 

  • OLED display.

 

  • Function of event marking.

 

  • Removable TF card(volume: up to 2GB).

 

  • One "AAA" size battery can collect the ECG data of 48-hour.

 

  • Ten electrodes synchronously collect the 12-lead ECG data.

The features of software

 

  •  With 12 leads synchro analysis , QRS search can be more exactly

 

  • There are more than 10 templates as atrial premature beat module, ventricular premature beat module, long interval module, atrial flutter module, atrial fibrillation module etc.and multiplicate User-defined templates could almost identify every kind of pathologic waveforms


 

  • Flexible atrial fibrillation analysis, so that physicians can use full, segmented automatic, manual analysis of atrial fibrillation, make the analysis more quickly and accurately

 

  • Powerful pacemaker analysis, on the AAI, VVI, DDD etc. all pacemaker analysis

 

  • Any period of single lead or 12-leads ECG can be reviewed at any time with the function of fast waveform review and analysis

 

  • Analysis of heart rate variability with short-range 5 minutes and long-range 1 hour and full analysis of heart rate variability

 

  • A one-stop printing operation, report printing process is convenient and quick

 

  • "Heart rate turbulence" analysis can predict the risk of death in patients with myocardial infarction

 

  • "T Wave alternation" analysis is a important index on judging and preventing Arrhythmia

 

  • The report has more reference value with the functions of VCG, VLP, TVCG and QTD analysis.


Main  Performance

Lead mode: standard 12-lead

Sampling accuracy: 12bit

Recording time: 48h

Scale voltage: 1mV±5%

Standard Sensitivity: 10mm/mV±5%

Noise level: ≤30μV

CMRR: ≥60dB

Low-frequency characteristics: Time constant≥3.2s

Enduring polarization voltage:±300mV DC polarization voltage,sensitivity shift≤±10%

Least measure signal: 50 µV p-p

Product safety type: Type B applied part(Internally powered equipment)

Dimension: 80.5mm(L) *59.5mm(W) *22mm (H)

Weight: 61.5g (Without battery)   Packed Weight: about 530.00G


bigPhoto
skype Ms. Jane

Tel: 86-28-87652638
Fax: 86-28-87652638
Mobile: 139 8203 8085
Contact to this supplier

Didn't find what you're looking for? Post Buying Lead or contact HiSupplier Customer Service Center  for help!

Related Search

Find more related products in following catalogs on Hisupplier.com